EasyManua.ls Logo

Teledyne 200E - Cleaning the Reaction Cell; Figure 9-5: no 2 Converter Assembly

Teledyne 200E
302 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model 200E Instruction Manual Instrument Maintenance
Remove the tube fittings from the converter.
Disconnect the power and the thermocouple of the converter. Unscrew the
grounding clamp of the power leads with a Phillips-head screw driver.
Figure 9-5: NO
2
Converter Assembly
Remove the converter assembly (cartridge and band heater) from the can. Make a
note of the orientation of the tubes relative to the heater cartridge.
Unscrew the band heater and loosen it, take out the old converter cartridge.
Wrap the band heater around the new replacement cartridge and tighten the screws
using a high-temperature anti-seize agent such as copper paste. Make sure to use
proper alignment of the heater with respect to the converter tubes.
Replace the converter assembly, route the cables through the holes in the can and
reconnect them properly. Reconnect the grounding clamp around the heater leads
for safe operation.
Re-attach the tube fittings to the converter and replace the insulation and cover.
Replace the instrument cover and power up the analyzer.
Allow the converter to burn-in for 24 hours, then re-calibrate the instrument.
9.3.9. Cleaning the Reaction Cell
The reaction cell should be cleaned whenever troubleshooting suggests. A dirty reaction cell
will cause excessive noise, drifting zero or span values, low response or a combination of
044100102 Rev A 151

Table of Contents

Related product manuals