EasyManua.ls Logo

Teledyne 200E - Figure 3-8: Status Output Connector; Table 3-2: Analog Output Pin-Outs; Table 3-3: Status Output Signals

Teledyne 200E
302 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Getting Started Model 200E Instruction Manual
Table 3-2: Analog Output Pin-Outs
Pin Analog Output Data Type Voltage Signal Current Signal
1 V Out I Out +
2
A1 NO
X
Ground I Out -
3 V Out I Out +
4
A2 NO
Ground I Out -
5 V Out I Out +
6
A3 NO
2
Ground I Out -
7 V Out not available
8
A4 Test Channel
Ground not available
If you wish to utilize the analyzer’s status outputs to interface with a device that accepts
logic-level digital inputs, such as programmable logic controller (PLC) chips, you can access
them through a 12 pin connector on the analyzer’s rear panel labeled STATUS.
EMITTER BUS
FOR PINS 1-8
STATUS
1 2 3 4 5 6 7 8 D
+
SYSTEM O
K
HIGH RANGE
CONC VALID
ZERO CAL
SPAN CAL
DIAGNOSTIC
MODE
LOW SPAN
Figure 3-8: Status Output Connector
Table 3-3: Status Output Signals
Pin # Status Condition (ON = Conducting)
1 SYSTEM OK ON if no faults are present.
2 CONC VALID ON if concentration measurement (NO, NO2 or NOx) is valid.
OFF any time the hold-off feature is active.
3 HIGH RANGE ON if unit is in high range of the Auto Range Mode.
4 ZERO CAL ON whenever the instrument is in ZERO point calibration mode.
5 SPAN CAL ON whenever the instrument is in SPAN point calibration mode.
6 DIAG MODE ON whenever the instrument is in diagnostic mode.
7 Unused
8 Unused
D EMITTER BUS The emitters of the transistors on pins 1-8 are tied together.
Unused
+ DC POWER + 5 VDC, 300 mA (combined rating with Control Output, if used).
Digital
Ground
The ground level from the analyzer’s internal DC power supplies
18 044100102 Rev A

Table of Contents

Related product manuals