Calibration Procedures Model 200E Instruction Manual
To program the sample sequence shown above, follow this flow chart:
SAMPLE RANGE = 500.0 PPB NOX=X.X
< TST TST > CAL CALZ CZLS SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG ACAL DAS RNGE PASS CLK MORE EXIT
SETUP X.X SEQ 1) DISABLED
NEXT MODE EXIT
SETUP X.X SEQ 2) DISABLED
PREV NEXT MODE EXIT
SETUP X.X MODE: DISABLED
NEXT ENTR EXIT
SETUP X.X MODE: ZERO
PREV NEXT ENTR EXIT
SETUP X.X MODE: ZERO
–
SPAN
PREV NEXT ENTR EXIT
SETUP X.X SEQ 2) ZERO
–
SPAN, 1:00:00
PREV NEXT MODE SET EXIT
SETUP X.X TIMER ENABLE: ON
SET> EDIT EXIT
SETUP X.X STARTING DATE: 01
–
JAN
–
02
<SET SET> EDIT EXIT
SETUP X.X STARTING DATE: 01–JAN–02
0
4 SEP
0
3
ENTR
EXIT
Toggle keys
to set
day, month &
year:
Format :
DD-MON-Y
Y
SETUP X.X STARTING DATE: 04
–
SEP
–
03
<SET SET> EDIT EXIT
Default
value is
ON
SETUP C.4 STARTING DATE: 04
–
SEP
–
03
<SET SET> EDIT EXIT
SETUP C.4 STARTING TIME:00:00
<SET SET> EDIT EXIT
SETUP C.4 STARTING TIME:00:00
1 4 : 1 5 ENTR EXIT
Toggle keys to
set time:
Format : HH:MM
This is a 24 hr
clock .
PM hours are
13 – 24.
Example
2:15 PM = 14:15
Toggle keys
to set
number of
days
between
procedures
(1-367)
Toggle keys
to set
delay time for
each iteration
of the
sequence:
HH:MM
(0 – 24:00)
SETUP C.4 DELTA DAYS: 1
<SET SET> EDIT EXIT
SETUP C.4 DELTA DAYS: 1
0 0 2 ENTR EXIT
SETUP C.4 DELTA DAYS:2
<SET SET> EDIT EXIT
SETUP C.4 DELTA TIME00:00
<SET SET> EDIT EXIT
SETUP C.4 DELTA TIME: 00:00
0 0 :3 0 ENTR EXIT
SETUP C.4 DELTA TIEM:00:30
<SET SET> EDIT EXIT
SETUP C.4 STARTING TIME:14:15
<SET SET> EDIT EXIT
EXIT
returns
to the
SETUP
Menu
SETUP C.4 DURATION:15.0 MINUTES
<SET SET> EDIT EXIT
SETUP C.4 DURATION 15.0MINUTES
3 0 .0 ENTR EXIT
SETUP C.4 DURATION:30.0 MINUTES
<SET SET> EDIT EXIT
Toggle keys
to set
duration for
each
iteration of
the
sequence:
Set in
Decimal
minutes
from
0.1 – 60.0
SETUP C.4 CALIBRATE: OFF
<SET SET> EDIT EXIT
SETUP C.4 CALIBRATE: OFF
ON ENTR EXIT
SETUP C.4 CALIBRATE: ON
<SET SET> EDIT EXIT
Toggle key
between
Off and
ON
SETUP C.4 SEQ 2) ZERO
–
SPAN, 2:00:30
PREV NEXT MODE SET EXIT
Sequence Delta Time
MODE Delta Days
7.9. Calibration Quality Analysis
After completing one of the calibration procedures described above, it is important to
evaluate the analyzer’s calibration SLOPE and OFFSET parameters. These values describe
the linear response curve of the analyzer, separately for NO and NO
X
. The values for these
terms, both individually and relative to each other, indicate the quality of the calibration. To
124 044100102 Rev A