EasyManua.ls Logo

Teledyne 200E - Control Inputs; Figure 6-4: Status Output Connector; Table 6-14: Status Output Pin Assignments

Teledyne 200E
302 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model 200E Instruction Manual Operating Instructions
The status outputs are accessed through a 12 pin connector on the analyzer’s rear panel
labeled STATUS. The function of each pin is defined in Table 6-14.
COMMON
EMITTERS
STATUS
1 2 3 4 5 6 7 8 D
+
SYSTEM O
K
HIGH RANGE
CONC VALID
ZERO CAL
SPAN CAL
DIAGNOSTIC MODE
LOW SPAN
GROUND
Figure 6-4: Status Output Connector
Table 6-14: Status Output Pin Assignments
Connector
Pin
Status Condition (ON=conducting)
1 SYSTEM OK ON if no faults are present.
2 CONC VALID ON if concentration measurement is valid, OFF when invalid.
3 HIGH RANGE ON if unit is in high range of any AUTO range mode.
4 ZERO CAL ON whenever the instrument is in ZERO calibration mode.
5 SPAN CAL ON whenever the instrument is in SPAN calibration mode.
6 DIAG MODE ON whenever the instrument is in DIAGNOSTIC mode.
7 LOW RANGE ON if unit is in low range of any AUTO range mode.
8 Unused.
D EMITTER BUS
The emitters of the transistors on pins 1-8 are bussed
together. For most applications, this pin should be connected
to the circuit ground of the receiving device.
+ DC POWER
+ 5 VDC, 30 mA maximum (combined rating with Control
Inputs).
DIGITAL
GROUND
The ground from the analyzer’s internal, 5/±15 VDC power
supply.
6.8.2. Control Inputs
Control inputs allow the user to remotely initiate ZERO and SPAN calibration modes are
provided through a 10-pin connector labeled CONTROL IN on the analyzer’s rear panel.
These are opto-isolated, digital inputs that are activated when a 5 VDC signal from the “U”
pin is connected to the respective input pin.
044100102 Rev A 73

Table of Contents

Related product manuals