Model 200E Instruction Manual APPENDIX A
Name Description Units
NOZSC2 NO concentration for range #2 during zero/span calibration,
just before computing new slope and offset
ppb
2
N2ZSC1 NO
2
concentration for range #1 during zero/span calibration,
just before computing new slope and offset
ppb
2
N2ZSC2 NO
2
concentration for range #2 during zero/span calibration,
just before computing new slope and offset
ppb
2
NXCNC1 NO
X
concentration for range #1 ppb
2
NXCNC2 NO
X
concentration for range #2 ppb
2
NOCNC1 NO concentration for range #1 ppb
2
NOCNC2 NO concentration for range #2 ppb
2
N2CNC1 NO
2
concentration for range #1 ppb
2
N2CNC2 NO
2
concentration for range #2 ppb
2
STABIL Concentration stability ppb
2
AZERO Auto zero offset (range de-normalized) mV
O3FLOW Ozone flow rate cm³/min
RCPRES Reaction cell pressure in-Hg-A
RCTEMP Reaction cell temperature °C
MFTEMP
1
Bypass or dilution manifold temperature °C
IZTEMP
1
IZS block temperature °C
CNVEF1 Converter efficiency factor for range #1 —
CNVEF2 Converter efficiency factor for range #2 —
CNVTMP Converter temperature °C
PMTTMP PMT temperature °C
SMPFLW Sample flow rate cm³/min
SMPPRS Sample pressure in-Hg-A
BOXTMP Internal box temperature °C
HVPS High voltage power supply output V
REFGND Ground reference (REF_GND) mV
RF4096 4096 mV reference (REF_4096_MV) mV
TEST11 Diagnostic test input (TEST_INPUT_11) mV
TEST13 Diagnostic test input (TEST_INPUT_13) mV
TEMP5 Diagnostic temperature input (TEMP_INPUT_5) °C
TEMP6 Diagnostic temperature input (TEMP_INPUT_6) °C
1
Factory option.
2
Standard instrument units.