SYSTEM OPERATION
To Review Live Data
3-48 TP04300 Series Interface & Applications Manual
To Review Live Data
Procedure 1. Begin controlling to temperature(s):
run to a setpoint from either the Operator or Cycle Screen.
2. Press HISTORY tab to review Live data
3. Press "Update From Live Data"
4. Press "Autoscale Time"
5. Press "Autoscale Temperature"
use Sliders, and Slider Fine Move, to adjust History Graph display
ATTENTION
"Live" data is unsaved data: when live data memory space full, earliest data is progressively
overwritten; if a new Test Setup is run, it progressively overwrites the prior session data (first
in data, is first out of the memory).