ROUTINE MAINTENANCE
RTD Sensor Calibration (DUT Mode)
5-32 TP04300 Series Interface & Applications Manual
RTD Sensor Calibration (DUT Mode)
RTD Sensor
Calibration
Introduction
There are 2 methods for calibrating an RTD Sensor.
Calibrator
External Sensor
Calibrator is the recommended method. This method uses an external voltage source (or
Calibrator) as the primary reference. The Calibrator inputs millivolt values for high and low set
points.
External Sensor is used for unique applications. It is recommended that you contact a
Temptronic representative before calibrating with this method.
This method uses an external temperature sensor as a primary reference.
Once the RTD Sensor has been chosen from the Calibration Select Sensor Screen, the
following dialog box will appear:
1980_507.jpg
Calibrator Method
(RTD Sensor)