REMOTE INTERFACES
IEEE Mandatory Commands
TP04300 Series Interface & Applications Manual 4-15
*STB? Read the status byte.
bit 7 - ready
bit 6 - master summary status (MSS) bit
bit 5 - standard event status (ESB) summary bit
bit 4 - message available (MAV) (GPIB only, always 0 for RS-232)
bit 3 - temperature event (TESR) summary bit
bit 2 - device specific error (EROR) summary bit
bit 1 - not used (always 0)
bit 0 - not used (always 0)
*TST? Self test (dummy, always returns 0, meaning “passed”).
*OPC Not implemented
*OPC? Not implemented
*WAI Not implemented
Command Description