CAUTION 9: When making the system air connection to the System, hold the AIR INPUT
fitting with a second wrench while tightening the barb fitting to prevent the AIR INPUT fitting
from rotating in the panel.
CAUTION 10: Use suitable Clean Dry Air (CDA) compressed air supply for the System: a) to
prevent premature fouling of the filters/regulator assemblies provided with the System, b) to
prevent ice forming from within the cooling module and possibly reducing or obstructing
output air flow. Improper air supply quality can cause damage to System internal operating
components.
CAUTION 11: Properly use and maintain the provided filters/regulator assemblies. Doing so
prevents moisture and/or compressor oils from being introduced into System operating
components. If left unchecked, moisture and/or compressor oil can cause damages to the
System which are not covered under warranty.
CAUTION 12: To loosen the linear actuator assembly (approximately 3 feet long), first
elevate the system high enough to allow the actuator to drop down clear (use a fork lift truck).
Then loosen the actuator from the rear of the system. Do not position yourself beneath (it is not
necessary to be under) the elevated system.