ROUTINE MAINTENANCE
TC Meter Calibration (TC Meter Mode)
TP04300 Series Interface & Applications Manual 5-47
Step Action
1 Press the Change Calibration Name button to re-name the calibration file as
desired.
If you do NOT re-name the Calibration file, you will overwrite the current
Calibration file.
2 Set the Ambient Window, C as desired.
Ambient is set at 25.0 C (+/- 20.0)
The Air temperature and the Calibrated DUT temperature will be the same
value when the system is within the Ambient Window.
3In the Low Calibration Point, C field, enter the desired “Low Calibration”
setpoint (between -60.0 and 0.0).
4In the Low DUT Calibrated Temperature, C field, enter the value of the Low
Calibration Point + the offset.
The offset must be a known (pre-determined value).
In the Example above, the Low Calibration Point, C is -60.0, the offset is -5.0
and the Low DUT Calibrated Temperature, C is -55.0.
5In the High Calibration Point, C field, enter the desired “High Calibration”
setpoint (between 40.0 and 200.0).
6In the High DUT Calibrated Temperature, C field, enter the value of the
High Calibration Point + the offset.
The offset must be a known (pre-determined value).
In the Example above, the High Calibration Point, C is 150.0, the offset is
-5.0 and the High DUT Calibrated Temperature, C is -145.0.
7 Press Save Calibration to save the file.
8 Press Done to exit the calibration screen