4
Note: US Patent no. 4,734,872
Temperature Range:
1,3
-80° to +225°C (60 Hz System)
-75° to +225°C (50 Hz System)
Typical Temperature Transition Rate (air)
1,2
-55° to +125°C: approx. 10 seconds or less
2
+125° to -55°C: approx. 10 seconds or less
2
System air flow output 2.4 l/s to 9 l/s (5 to 18 scfm) CONTINUOUS
Temperature accuracy 1.0°C (when calibrated against NIST transfer standard)
Temperature set, display and resolution ± 0.1°C
Temperature Control:
DUT Sensor Ports Internal diode, Type T and Type K Thermocouple and
100 ohm Platinum RTD
DUT Control Control to within ± 0.1°C; SELF-TUNING available in DUT Control
Remote interface ports IEEE-488, RS232C Serial, and Start Test/End of Test/Stop on
First Fail (ST/ET/SFF) and Ethernet
1
Note: Ultimate low temperature and system performance may vary under operating conditions less than or greater than nominal.
2
Note: Transition performed under nominal operating conditions.
3
Note: Low temperature extremes may be achieved at reduced airflow rates.
TP04300B
Model TP04300B ThermoStream
®
System Specifications
USER FEATURES
P
art No.
SL10420 R/A
Mobile Temperature System for testing components, parts, hybrids, modules,
subassemblies and printed circuit boards at precise temperature.
TEMPERATURE PERFORMANCE AND
AIRFLOW CAPACITY
System Configuration Options (Choice):
Modes of operation Two: Operator Mode and Cycling Mode
Test set-up configurations In Cycling Mode, an unlimited quantity
may be created; save to hard disk/diskettes
Ramp/soak/cycle configurations
In Cycling Mode, up to 18 sequences per test
set-up; table is displayed on-screen
Program and data storage Data logging and program files may be stored
on the hard drive or to a 3.5" floppy diskette
External Device Ports Ports are located on the front of the system for
connecting a mouse, printer, keyboard; plus USB Port
(for USB memory stick, USB mass storage drive, USB
printer
, etc.).
On-screen help Included for both Cycling and Operator’s Modes
DUT temperatur
e contr
ol
Patented
4
Dual Loop T
emperatur
e Contr
ol
Status indicators
On-scr
een and r
emote I/O
Purge flow tester interface Dry air purge protects tester electronics
protection from condensation
Purge flow capacity 0.25 to 1.5 liters per second (0.5 to 3 scfm) airflow,
manually adjustable
Calibration
Automated, simplified and accurate for all
airflows and DUT types
Range of motion Turret assembly (connecting Flexible Extender Hose)
rotates 360º for full freedom of positioning of Flexible
Extender Hose ar
ound base unit
Turret Assembly Rotation: 360°
of fr
eedom ar
ound system base.
• Modular TP04300B base unit with Turret Assembly and Flexible Extender Air Transfer Hose
(Specify choice of Flexible Extender length -up to 6 ft. is included with system. Longer Flexible Extender Hoses are also available- contact factory.)
•
Modular TP04300B base unit with Adapter Kit for connecting a ThermoChamber
TM
Accessory (ThermoChamber enclosure is sold separately.)
TP04300B with
optional ThermoFixture
TP04300B with
optional ThermoChamber