Appendix B RS-232 Commands
B-8
set range
d
set low range
d
d = Code in table below
The “set range” command selects the CO fullscale range according to the table below for
single mode and sets the high range for dual and auto range. The “set low range”
command sets the low range in dual or auto range. The example below sets the CO
fullscale range to 50 ppm.
Send: "set range 5"
Receive: "set range 5 ok"
Code ppm mg/m
3
0 1 1
1 2 2
2 5 5
3 10 10
4 20 20
5 50 50
6 100 100
7 200 200
8 500 500
9 1,000 1,000
10 2,000 2,000
11 5,000 5,000
12 10,000 10,000
13 C1 C1
14 C2 C2
15 C3 C3