Appendix B RS-232 Commands
B-9
custom
d
d = | 1 | 2 | 3 |
This reports the user-defined value of custom range 1, 2, or 3. The example below
reports that custom range 1 is defined to 55.0 ppm.
Send: "custom 1"
Receive: "custom 1 5500E-2 ppm"
set custom 1 range
ddddd.d
set custom 2 range
ddddd.d
set custom 3 range
ddddd.d
These commands are used to define the custom ranges. To use the custom range select it
using the set range command. The example below defines custom range 1 to 55.5 ppm.
Send: "set custom 1 range 55.5"
Receive: "set custom 1 range 55.5 ok"
avg time
low avg time
The “avg time” command reports the averaging time setting in single mode or the high
averaging time in dual or auto range. The “low avg time” command reports the low
averaging time in dual or auto range. The example below reports that the averaging time
is set to 60 seconds.
Send: "avg time"
Receive: "avg time 060 sec"