Appendix B RS-232 Commands
B-10
set avg time
d
set low avg time
d
d = Code in table below
The “set avg time” command sets the averaging time according to the Table below for
single mode and sets the high averaging time in dual or auto mode. The “set low avg
time” sets the averaging time for the low CO in dual or auto range. The example below
sets the averaging time to 60 seconds.
Send: "set avg time 3"
Receive: "set avg time 3 ok"
Code Averaging time (seconds)
0 1
1 2
2 5
3 10
4 20
5 30
6 60
7 90
8 120
9 180
10 240
11 300