Chapter 3 Operation
3-73
Default Coefficients The Default Coefficient Screen is used prior to a calibration
when there has been a bad calibration or new ranges have been set up.
To view the Default Coef Screen
From the Main Menu Select Service Mode
From the Service Mode Menu select Hi or Lo Multi Point Cal
From Hi or Lo Multi-Point Cal select Default Coef
Default Coef Screen
To use the Default Coef Screen
Press
ENTER
to set coefficients to their default values
The Analyzer will automatically reset the default values and save the default
parameters COEF 0= 1.000000, COEF 1= 0.000000, COEF 2= 0.000000.
COEF 0 1.000000
1 0.009000
2 0.000000
SET TO DEFAULT?