EasyManua.ls Logo

Thermo Scientific Applied Biosystems QuantStudio - Page 55

Thermo Scientific Applied Biosystems QuantStudio
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Observation Possible Cause Action
A bubble was present in the
loading blade when it was
used to apply the PCR
reaction to the chip.
If possible, use the AnalysisSuite
Software to lter the low quality data
points, or discard the chip and run the
sample again.
When loading the loading blades:
If you are using a manual pipette,
pipette to the rst stop.
Decrease your pipetting speed.
Observation Possible Cause Action
The chip leaked during
thermal cycling or
imaging.
A large bubble was
present in the chip
(insufcient Immersion
Fluid).
Excess Immersion Fluid
is present on the chip
lid.
If present, remove excess Immersion
Fluid from the chip lid and run the
chip again.
If possible, use the AnalysisSuite
Software to lter the low-quality data
points.
To minimize leakage:
Leave a small bubble when lling
chips with Immersion Fluid.
When covering chips with
Immersion Fluid, do not get uid
on the sides of the chip base.
Chapter 6 Troubleshooting
Troubleshooting chip images using the Chip View
6
QuantStudio
3D Digital PCR System User Guide
55

Table of Contents

Related product manuals