3.
Remove each fuse from its fuse holder and inspect it for damage. Carbon
typically coats the inside of failed fuses.
4.
Replace each failed fuse with a 1.6A, Time-Lag T, 250VAC, 5 × 20-mm Fuse.
Note: The voltage and amperage ratings are on the fuse holder.
5.
Install the fuse holder.
6.
Plug in, then power on the QuantStudio
3D Digital PCR Instrument. The
installation is successful if the instrument powers on.
Note: Fuse failure can result from uctuations in the supplied power to the
QuantStudio
3D Digital PCR Instrument. To prevent further failures, consider
installing an electrical protective device, such as a UPS or surge protector.
Calibrate the touchscreen
Calibrate the instrument touchscreen periodically as part of the routine maintenance
of your instrument, if you observe a decrease in screen responsiveness over time, or
if prompted after updating the instrument rmware.
1.
From the Main Menu of the touchscreen, touch
, then touch Maintenance &
Service.
2.
In the Maintenance & Service screen, touch Screen Calibration.