Network requirements
The QuantStudio
3D Digital PCR Instrument supports IPv4 TCP/IP communication
and provides two methods for integrating the device into a local area network (LAN,
also see “Materials required” on page 96):
The fast Ethernet adapter (10/100 Mbps) with a RJ45-type connector on the rear
panel of the instrument allows you to connect the device directly to a standard
network port using a standard Category 6 Ethernet cable of the required length.
An optional QuantStudio
3D Extended Arm WiFi Adapter (Cat. No. A28598)
connected to the instrument enables a connection to compatible wireless
hotspots.
The QuantStudio
3D Instrument requires the following if connected to a network:
The room where the instrument will be deployed must contain at least one active
network jack.
If the instrument will be located more than 10 feet from the network jack, you
must provide a shielded Category 6 Ethernet cable of sufcient length.
If dynamic network conguration is not available, a static IP address must be
reserved for the instrument.
The congured DNS server must resolve internet domain names (such as
"apps.lifetechnologies.com/" and "pool.ntp.org").
The names of any local le servers used as data destinations must also be
available through DNS, or optionally WINS.
If the instrument will transfer experiment (.eds) les directly to the
QuantStudio
3D AnalysisSuite
Software, then the instrument must have
outbound Internet access through port 443/HTTPS to
apps.lifetechnologies.com/quantstudio3d/.
If the instrument will use Network Time Protocol (NTP) for date/time
synchronization, the instrument must have outbound Internet access through
port 123/UDP to the server pool at pool.ntp.org.
If the instrument will transfer experiment (.eds) les to a Windows
operating
system-based or SAMBA le server, network credentials must be provided for
authentication.
In addition to the requirements above, if the QuantStudio
3D Instrument will be
connected to a wireless network, the access point must support:
IEEE 802.11b/g/n wireless standards
WEP, WEP-Enterprise, WPA- Personal, or WPA2-Personal security
Appendix B Specifications and Layout
Network requirements
B
QuantStudio
3D Digital PCR System User Guide
83