Thermal cycle the chips
Materials required .................................................... 32
Prepare the thermal cycler ............................................ 32
Guidelines for handling loaded chips ................................... 35
Perform the PCR ..................................................... 36
Unload the thermal cycler ............................................. 38
Materials required
Loaded and sealed QuantStudio
3D Digital PCR 20K Chip v2
ProFlex
2 × Flat PCR System (or Dual Flat Block GeneAmp
PCR System
9700)
QuantStudio
3D Digital PCR Chip Adapters
QuantStudio
3D Digital PCR Thermal Pads
QuantStudio
3D Digital PCR Tilt Base
Isopropanol
Gloves, powder-free, nitrile
Low-lint polyester wipes, clean-room grade
Prepare the thermal cycler
Before running each batch of chips, perform the following to ensure efcient thermal
cycling:
Conrm that clean QuantStudio
3D Digital PCR Thermal Pads are available for
use.
IMPORTANT! If a Thermal Pad becomes dirty, clean the pad using a low-lint
wipe sprayed with isopropanol, then let it dry. Fluids, debris, and other
contaminants that adhere to the pad can transfer to the chips during thermal
cycling and uoresce during imaging.
IMPORTANT! Use only Thermal Pads that are approved for use with chips.
Thermal compression pads from other PCR applications may not apply uniform
pressure or heat distribution to all chips on the sample block.
4
32
QuantStudio
3D Digital PCR System User Guide