Guidelines for handling loaded chips
To ensure the proper function of your loaded chips, handle them according to the
following guidelines:
Always wear powder-free gloves.
IMPORTANT! Oils from your hands can contaminate the chips and interfere
with imaging.
Handle the loaded chips by gently gripping the sides of each chip case, being
careful to not apply pressure to the chip within it.
Do not touch the window of the chip lid. If you accidentally touch the window,
clean it using a laboratory wipe that has been sprayed with isopropanol, then dry
by wiping in one direction using a clean low-lint wipe.
Note: Debris or excess sample on the lid can interfere with thermal uniformity
during thermal cycling, or uoresce during imaging.
Thermal cycle chips within 2 hours after loading them.