EasyManua.ls Logo

Thermo Scientific Applied Biosystems QuantStudio - Page 60

Thermo Scientific Applied Biosystems QuantStudio
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Observation Possible Cause Action
Too much sample was left in
loading blade.
Note: The top image on the
left was generated by
QuantStudio
3D Instrument
rmware v1.1 or earlier. The
bottom image was
generated by rmware v2.0
or later. In v2.0 or later, the
overloaded sections of the
chip appear in white and
these regions are ltered out
and removed from analysis.
Ensure that you are loading only
14.5 µL of reaction mix into the
loading blade.
Chapter 6 Troubleshooting
Troubleshooting chip images using the Chip View
6
60
QuantStudio
3D Digital PCR System User Guide

Table of Contents

Related product manuals