5-4 Error Messages, Diagnostic Tests, and Troubleshooting
1. Ensure that the flow cell is clean.
2. Ensure that the lamp access door is shut securely.
3. Recycle power to the detector.
4. If the catastrophic error persists, contact Waters Technical Service (see
page 4-2).
The following table has error messages that prevent operation, listed in
alphabetical order.
Error messages preventing operation
Error message Description Corrective action
Communication failure:
Reference A/D
A/D communication
test failed.
1. Power-off and
power-on again.
2. If the problem
persists, call Waters
Technical Service.
Communication failure:
Sample A/D
A/D communication
test failed.
1. Power-off and
power-on again.
2. If the problem
persists, call Waters
Technical Service.
Configuration not
found
Stored configuration
data is invalid.
Power-off and power-on
again.
Dark current too high:
nnnnnnn
The dark energy level is
above 1000000.
1. Power-off and
power-on again.
2. If the problem
persists, call Waters
Technical Service.
Dark current too low: 0 The dark energy level
equals 0.
1. Power-off and
power-on again.
2. If the problem
persists, call Waters
Technical Service.