EasyManua.ls Logo

Waters 2475 - Choosing Signal Connections

Waters 2475
228 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2-10 Setting Up the Detector
Tip: For pin-out connections to the 2475 detector, see the figure “I/O signal
inputs and outputs” on page 2-11.
Choosing signal connections
Tip: Connect the 2475 detector to other HPLC system components through an
Ethernet connection.
The rear panel provides two analog connectors and an RS-232
communications port for operating the detector with external devices. You can
connect other instruments to the detector through these connectors to enable
the following signals:
Analog outputs – Two attenuated, analog-channel outputs, Detector
Out 1 and Detector Out 2, support 1-V output to external devices or data
systems. For input/output voltage current specifications, see Appendix
B. The 1-V output for channel A and channel B is scaled according to the
EUFS (emission/energy units full scale) setting for each channel. The
detector allows the EUFS to be set individually for the output on each
channel. Volts per EU are calculated for the 1-V output as follows:
Volts out = Fluorescence × 1 V/EUFS
For example, an EUFS setting of 10,000 provides a traditional 0.0001
V/EU output. An EUFS setting of 100,000 provides a 0.00001 V/EU
output, which supports chromatography above 10,000 EU.
Chassis ground stud – Connect the shield from analog connections
here.
Switched outputs – You can program two switch contact closures to
turn on, off, toggle, pulse once for a defined duration, or pulse
repetitively for a specified period of time.
Event inputs – Four general-purpose TTL contact closures on the
detector’s A (Inputs) terminal support the following functions (see the
table titled “Primary and secondary function (method) parameters” on
page 3-21):
Remote or inject start
–Lamp on/off
Chart mark input
Auto Zero

Table of Contents

Related product manuals