EasyManua.ls Logo

Waters 2475 - Startup diagnostic tests; Mobile-phase solvent degassing

Waters 2475
228 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Startup diagnostic tests 1-21
The gain for Region 2 is determined by the maximum signal level in Region 1.
Therefore, a gain of only 10 would be used from time 0.0 to time 2.0, but the
small peak may not be adequately resolved at this setting. If the detector did
find it, the peak area integration would be far less accurate because of higher
baseline noise. Failure to program a gain change at a strategic point in the
chromatogram constitutes a relatively ineffective approach to method
development.
Startup diagnostic tests
The detector runs a series of startup diagnostic tests and posts an error
message if any test fails. The startup diagnostic tests are as follows:
Central processing unit (CPU) test
Serial communication interface (SCI) test
Electrically erasable programmable read-only memory (EEPROM) test
RAM test
Application program checksum verification
•Lamp test
•Photodiode test
•PMT test
Optics test/Wavelength verification
Mobile-phase solvent degassing
Mobile-phase difficulties account for at least 70% of all liquid chromatographic
problems. Using degassed solvents is important, especially at excitation
wavelengths shorter than 220 nm. Bubbles in the flow cell adversely affect
detector performance. Degassing provides
reproducible fluorescent response.
stable baselines and enhanced sensitivity.
reproducible retention times for eluting peaks.
reproducible injection volumes for quantitation.
stable pump operation.

Table of Contents

Related product manuals