Connecting other devices 2-23
Connecting other devices
You can connect many devices to the detector, including these:
Empower PC using the bus SAT/IN™ module
Millennium
32
chromatography workstation using the bus SAT/IN
module
Waters 746 Data Module
Chart recorder
Waters 600-series pump
Waters 717plus Autosampler
Waters Fraction Collector II or III
Required materials
To connect cables to the terminals on the detector’s rear panel, you need the
following tools:
Small flat-blade screwdriver (startup kit)
Electrical insulation stripping tool
Connecting cables
To connect the cables from other HPLC system devices to the A and B
terminals on the detector’s rear panel:
1. Remove terminal A or B (see the figure “I/O signal inputs and outputs”
on page 2-11).
2. Unscrew the connecting pin terminal.
3. Using the stripping tool, strip the wire about 1/8-inch from the end.
4. Insert the stripped wire into the appropriate connector.
5. Tighten the screw until the wire is held securely in place.
6. Reinsert the terminal.
7. Press the terminal firmly to ensure that it is inserted fully.