Troubleshooting 5-23
RS-232 problems Disabled RS-232
configuration
Set the Configuration
screen correctly.
Bad RS-232 cable Check and, if
necessary, replace the
RS-232 cable.
Xenon lamp does not
light
Faulty lamp Replace the lamp.
Lamp not plugged in Plug in the lamp
connector.
Bad lamp power supply Contact Waters
Technical Service.
Lamp switch off Check the rear panel
connections.
General hardware troubleshooting (Continued)
Symptom Possible cause Corrective action