3-22 Using the Detector
Analog out
(single λ)
2 (of 4) Choice None Emission
A
Reference
energy A
•Output off
Emission A
Analog out
(multi λλ)
2 (of 4) Choice None Emission
A
•MaxPlot A,
B, C, D
Diff (A-B)
Diff (B-A)
Reference
energy A
•Output off
Emission A
Time
constant
2 (of 4) Numeric sec Hamming
(λ): 0.1 to
5.0
Hamming
(λλ): 1 to
50
•RC(λ): 0.1
to 99
•RC(λλ): 1
to 99
•0 to
disable
filtering
1.5
Data units 3 (of 4) Choice None Emission
•Energy
Emission
Voltage offset 3 (of 4) Numeric mV Integer –1000
to +1000
0
Chart
polarity
3 (of 4) Choice None +
+
Primary and secondary function (method) parameters (Continued)
Function Screen Type Units Range Default