EasyManua.ls Logo

Waters 600e - Page 60

Default Icon
258 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
32 Preparing Your 600E System for Operation
Open the slide-out drawer. Disconnect the fluid tubing at the column inlet. See the
Waters 600E Multisolvent Delivery System Installation and Maintenance Guide,
Section 2.3.1, Installing a Column. Disconnecting the fluid line avoids introducing
air into the column and reduces backpressure resistance to priming.
If you are using a Waters 717plus Autosampler, put the autosampler in purge mode.
Initiating Flow
1. Rotate the inlet manifold valve handle located on the front of the pump to the Run
position (Figure 3-3).
2. Place a waste beaker under the vent tube of the reference valve to collect the
diverted eluent.
3. Open the reference valve (see Figure 1-4) by rotating the handle from left to right to
isolate the column from the flow path.
Figure 3-3 Pump Inlet Manifold Valve Positions
4. Press the Direct function key and the Direct Control screen appears. At the Flow
field, type a flow rate of 1.0 mL/min (the gradient proportioning valve is not
actuated at 0 mL/min).

Table of Contents

Other manuals for Waters 600e

Related product manuals