Use
28.03.2022 61 / 112
Abbreviation Program name Info on program
P12 Universal NoVac
2bar
Universal test for all types of watches that are classified as waterproof.
Corresponds to the 2bar test as per ISO 22810:2010-4.4 and ISO
6425:2018-B.1.1. Without vacuum test.
P13 Universal NoVac
3bar
For all types of watch that are classified as water-resistant to
3ATM/30m. Without vacuum test.
P14 Universal NoVac
5bar
For all types of watch that are classified as water-resistant to
5ATM/50m. Without vacuum test.
P15 Universal NoVac
10bar
For all types of watch that are classified as water-resistant to
10ATM/100m or more. Without vacuum test.
P99 Calibrated Leak For testing the function of the ProofMaster in combination with Witschi
reference leaks.
Parameters for the preset
programs
Abbreviation Program name Pressure values
(bar)
Limit values
(µg/min)
Watch size
(mm)
P01 Universal 1bar -0.2 / 1.0 200 / 50 20 - 40
P02 Universal 2bar -0.3 / 2.0 200 / 50
P03 Universal 3bar -0.4 / 3.0 200 / 50
P04 Universal 5bar -0.5 / 5.0 200 / 50
P05 Universal 10bar -0.7 / 1.0 / 10.0 200 / 200 / 50
P11 Universal NoVac
1bar
1.0 50
P12 Universal NoVac
2bar
2.0 50
P13 Universal NoVac
3bar
3.0 50
P14 Universal NoVac
5bar
5.0 50
P15 Universal NoVac
10bar
1.0 / 10.0 50
P99 Calibrated Leak 2.0 50
Info
The 50 µg/min limit is only applied at the highest test
pressure of a program, as per ISO 22810. The previous test
pressures are tested for the preliminary test with a limit of
200 µg/min.