Settings
84 / 112 28.03.2022
ð Device is highlighted in blue and the right window displays
the device options.
b) Activate or deactivate the function Start by pressing the bell
cover.
ð The button is blue when the function is enabled and grey
when it is not.
7.4.3 Sensor lighting
a) Select Device from the list.
ð Device is highlighted in blue and the right window displays
the device options.
b) Enable or disable the sensor lighting.
ð The button is blue when the function is enabled and grey
when it is not.
7.4.4 Quiet pressure release
This setting allows the test pressure to be released more slowly from
the test chamber.
This reduces the noise emission but increases the test time.
For a test at 10bar, the sound level is reduced from 75dB(A) to
68dB(A) and the discharge can be increased by 15seconds.