inoLab
®
Multi 9620 IDS Dissolved oxygen
ba77173e04 12/2017 79
Calibration
evaluation
FDO
®
925
Calibration record
Display Calibration record Relative slope
+++ S = 0.94 ... 1.06
++ S = 0.92 ... 0.94
or
S = 1.06 ... 1.08
+ S = 0.90 ... 0.92
or
S = 1.08 ... 1.10
Error Error S < 0.90
or
S > 1.10
Error elimination (see section 15 W
HAT TO DO IF...,
page 125)
CALIBRATION Ox
01.09.2017 07:43:33
FDO
®
925
Ser. no. 10146858
SC-FDO 925 10158765
Relative slope 0.98
Sensor +++