<3.  INSTALLATION>
3-4
IM 01E30D01-01EN
(4)  Terminal Box Cover
IMPORTANT
As it is possible that the insulation will 
deteriorate, do not open the terminal box cover 
until it is time to wire it.
(5)  Long-term Non-use
IMPORTANT
Itisnotdesirabletoleavetheowmeterunused
for a long term after installation. If this situation 
isunavoidable,takecareoftheowmeterby
observing the following.
• Conrmationofsealingconditionsforthe
owmeter
Conrmthattheterminalboxscrewandwiringports
are well sealed. Equip the conduit piping with drain 
plugs or waterproof glands to prevent moisture or 
waterfrompenetratingintotheowmeterthrough
the conduit.
• Regularinspections
Inspect the sealing conditions as mentioned above, 
and the inside of the terminal box at least once a 
year. Also, due to rain, etc. when it is suspected 
that water may have penetrated into the inside 
owmeterperformsupplementaryinspections.
3.2.2   Flowmeter Piping
CAUTION
Misaligned or slanted piping can lead to leakage 
anddamagetotheanges.
(1)  Correct any misaligned or slanted piping, and 
any gaps that may exist between mounting 
angesbeforeinstallingtheowmeter(Read
Figure 3.2.1).
Slanted
Misaligned
Figure 3.2.1  Slanted and Misaligned Flowmeter 
Piping
(2)  Inside a newly installed pipeline, there may 
be some foreign substances such as residue 
from welding or wood chips. Remove them 
byushingthepipingbeforemountingthe
owmeter.Thispreventstheliningfrom
being damaged, as well as the occurrence 
of erroneous measured signals resulting 
from foreign substances passing through the 
owtubeduringmeasurement.