<11.  MAINTENANCE>
11-6
IM 01E30D01-01EN
11.4  Adhesion diagnostic 
function
Thelong-termmeasurementofadhesiveuids
mayadverselyaffecttheowmeasurementdue
to insulating materials adhering to the electrode, 
which could cause the output vibration to increase 
ormakeitimpossibletomeasuretheowrate.
This function checks the level of adhesion by 
measuring the resistance value.
NOTE
Adhesion Diagnostic Function
• Thisfunctiondiagnoseadhesionusing
electrode resistance values.
• Theresultofadhesiondiagnosisisdisplayed
at K12:Adh Measure Value.
• When“Adhesioncheck”hasbeensetfor
B41/H11:DisplaySelect2 or B42/H12:
Display Select 3, the diagnose adhesion 
is indicated on the display unit using four 
different levels.
• IfthejudgmentvalueforLevel3is
exceeded, a warning is displayed; and if 
the value for Level 4 is exceeded, an alarm 
or warning is displayed depending on the 
setting condition at G29.
• Availableconductivityforthisfunctionis
limited to 10 µS/cm.
  Make sure to use the adhesion diagnostic 
function with the greater conductivity than 
the above mentioned value. 
IMPORTANT
• Parameterscannotbesetduringexecution
(approximately 5 minutes).
• Adhesiondiagnosisshouldonlybecarried
outwhentheuidvelocityiscompletelyzero
by closing the valve.
• Flowmeasurementisnotperformedduring
execution. While adhesion diagnosis is being 
carried out, normally a current of 4 mA is 
output. When adhesion diagnosis is carried 
out, change the control loop to the manual 
moderst.
• Adhesiondiagnosiscannotbecarriedout
while an alarm is occurring.