<9.  OPERATION VIA HART CONFIGURATION TOOL (HART 7)>
9-5
IM 01E30D01-01EN
• Procedureofdevicevariablesimulation
Step 1 Call up the 
parameter 
[Root Menu]→Diagnosis/
Service→Test→Simulate
2 Selection of 
Device Variable
Select one parameter from 
the list below
 Off
PV(Instantanousow
rate)
 SV (Forward totalization 
value)
 TV (Reverse totalization 
value)
 QV (Differential totalization 
value)
 % Range (% in output)
 Loop Current (current 
output)
3 Setting of Value Input the simulate value
4 Setting of Data 
quality
Select one parameter from 
the list below
 Bad
 Poor accuracy
 Manual / Fixed
 Good
5 Setting of Limit 
status
Select one parameter from 
the list below
 Not limited
 Low limited
 High limited
 Constant
NOTE
• The total is accumulated from the 
instantanousowrate.Therefore,thetotal
dependsontheinstantanousowrate.
• The total simulation is only applied for LCD 
display and communication output and does 
not affect the totalization value.
• Thesimulationsforowratearereected
to the output. Accordingly, the loop current, 
LCD display, and communication output are 
directly corresponded to the simulate value. 
The alarm output is also available according 
to the simulate value.
(3) Squawk 
This feature can be used to identify the 
communicating device by remotely causing LCD 
to display the particular pattern as shown in the 
Figure.
“SQUAWK” continues for approximately 60 
seconds, then is released automatically.
• ProceduretocalluptheSquawk display
[Root Menu]→Diagnosis/Service→Test→Squawk
→Squawk
9.8.2  Trim Analog Output
This function is used to adjust the analog output at 4 
mA and 20 mA with the D/A trim.
(1)  D/A trim
  Connect a calibration digital ammeter, and then 
enter the read value of the ammeter for each 
outputofAXR.
• ProceduretocalluptheD/Atrim display
[Root Menu]Diagnosis/Service→Adjustment→D/A
trim
(2)  Clear D/A trim
  When output is reset to the original value, clear 
the D/A trim.
• ProceduretocalluptheClearD/Atrim display
[Root Menu]Diagnosis/Service→Adjustment→Clear
D/A trim
CAUTION
The output adjustment function can match the 
4mA and 20mA output to the reference meter 
such as a voltmeter. In the output adjustment, it 
is necessary to use the calibrated voltmeter and 
resistance.