<13. PED (PRESSURE EQUIPMENT DIRECTIVE)>
13-1
IM 01E30D01-01EN
This chapter describes further requirements and
notices concerning the PED (Pressure Equipment
Directive). The description in this chapter is prior to
other description in this User’s Manual.
(1) Technical Data
Module: H
Type of Equipment: Piping
Type of Fluid: Liquids and gas
Group of Fluid: 1 and 2 (*4)
General-purposeUse/ExplosionProofType:
MODEL
DN
(mm)
(*1)
PS
(MPa)
(*1)
PS · DN
(MPa ·
mm)
CATEGORY
(*2) (*4)
AXR025G
25 4 100
Sound
Engineering
Practice
(SEP) (*3)
AXR025C
AXR040G
40 4 160 II
AXR040C
AXR050G
50 4 200 II
AXR050C
AXR065G
65 2 130 II
AXR065C
AXR080G
80 2 160 II
AXR080C
AXR100G
100 2 200 II
AXR100C
AXR150G
150 2 300 II
AXR150C
AXR200G
200 2 400 III
AXR200C
*1: PS: Maximum allowable pressure for Flowtube
DN: Nominal size
*2: For details, refer to the following.
Table6coveredbyANNEXIIofDirective97/23/EC
(until July 18th, 2016)
Table6coveredbyANNEXIIofDirective2014/68/EU
(from July 19th, 2016)
*3: Article 3, paragraph 3 of Directive 97/23/EC
(until July 18th, 2016)
Article 4, paragraph 3 of Directive 2014/68/EU
(from July 19th, 2016)
*4: Modelsclassiedin
CATEGORY II shall not be used for
unstable gases of Group 1.
(2) Installation
WARNING
• Tightentheboltsofthepipingjoints
according to the prescribed torque values.
• Takemeasurestoprotecttheowmeters
from forces caused by vibration channeled
through the piping.
(3) Operation
WARNING
• Theinstrumentshouldbeoperatedwiththe
temperatureandpressureoftheuidunder
normal operating conditions.
• Theambienttemperatureshouldbethatof
normal operating conditions.
• Takemeasurestopreventexcessive
pressure such as water hammer, etc. To
avoid water hammer prevent the pressure
from exceeding the PS (maximum allowable
pressure) by setting the system’s safety
valves, etc. appropriately.
• Shouldexternalreoccur,takesafety
measures at the device itself or system-wide
preventithavinganeffectontheowmeters.
• Avoidusinguidsexceedingthecorrosion
proof limitations of the lining and electrodes.
• Takemeasuresnottoabradethemetalpipe,
andavoidabradingtheliningbyusinguids
such as slurry and sand are contained.
13.
PED (PRESSURE EQUIPMENT DIRECTIVE)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA ADMAG AXR and is the answer not in the manual?

YOKOGAWA ADMAG AXR Specifications

General IconGeneral
BrandYOKOGAWA
ModelADMAG AXR
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals