<6.  PARAMETER DESCRIPTION>
6-44
IM 01E30D01-01EN
6.5.3   Alarms & Warning Messages
System Alarms (Device breakdown or inability to obtain correct measurements.)
Displayunit/BRAIN 
(60) content
Alarm countermeasure 
message on display unit
NE-107 
status
Alarm description Countermeasure
10:μPFault Contactnearofceor
service center
F Microprocessor (CPU) failure Contact your nearest 
Yokogawaofceorservice
center
11:EEPROM Fault F EEPROM failure
12:Sub uP Fault F Substitute mcroprocessor (CPU) failure
13:EXPwrFault F Extitation power circuit failure
14:A/D(S) Fault F A/D converter failure
15:A/D(I) Fault
16:Analog Fault F Analog circuit failure
17:Coil Open F Flowtube coil is open-circuit
18:Coil Short F Flowtube coil is short-circuit
19:Excite Error F Excitation circuit failure 
20:Pulse Error F Discrepancy of pulse count
21:EEPROMDt F EEPROM default values
Process Alarms (Device is normal but process-related errors make correct measurement impossible.) 
Displayunit/BRAIN 
(60) content 
Alarm countermeasure 
message on display unit 
NE-107 
status
Alarm description  Countermeasure 
30:SigOverow Check signal cableand 
grounding 
S Input signal error  Carry out an investigation as follows: 
•Checkthesignalcableforbreakage.
•Checkforstraycurrentsintheuid.
•Check the grounding. 
•Checkwhetherowtubeisllingwithuid.
•Checktheelectrodeadhesion.
31:Empty Pipe  Fillowtubewithuid S Flowtubeisnotlledwithuid Filltheowtubewithuid.
32:H/L Alm  Checktheowrateand
setting value 
–
Flow rate alarm for greater 
than High limit value or less 
than Low limit value. 
Checktheowrateandsettingvalueof
High limit and Low limit. 
33:Adhesion Alm  Clean electrodes  S Electrode adhesion alarm  Clean the electrodes.