<6. PARAMETER DESCRIPTION>
6-16
IM 01E30D01-01EN
6.4 Parameter Description
IMPORTANT
One output can be selected from pulse, alarm, or
status through the parameter setting.
(1) Menu B: Easy Setup items
Those parameters with a high frequency of use
have been grouped together in Easy Setup. All
basic functions can be controlled using only the
parameters from this block. Parameters from
Menu B share identical names with those from
othermenus;however,modicationofonesuch
parameter will result in the other being automatically
modied.
[B10: Language] Selection of language used for
the display unit
This setting is linked with that of parameter H30.
One of the following languages can be selected for
the display unit.
Data Range
Setting item Description
English All parameters, alarm messages, etc. displayed
in English.
Japanese All parameters, alarm messages, etc. displayed
in Japanese katakana.
French All parameters, alarm messages, etc. displayed
in French.
German All parameters, alarm messages, etc. displayed
in German.
Italian All parameters, alarm messages, etc. displayed
in Italian.
Spanish All parameters, alarm messages, etc. displayed
in Spanish.
[B20: Flow Damping] Setting of the damping time
constant
This setting is linked with that of parameter C11.
Thedampingtimeconstantshouldbemodiedto
suppressanoutputuctuationortochangethe
response time. This time constant has an effect
onanalogoutputandontheowratedisplay(i.e.,
actualinstantaneousowrate,%,currentvalue),
and in addition, it also affects pulse output and
totalization. However, when “No Damping” has
been set for J25:T/PDampSelect, there will be no
effect on pulse output or totalization.
*: Time constant: The time required for the output to reach
63.2% from 0%.
NOTE
When the damping time is short, the output
uctuates.Setthetimeto5secondsormorefor
control loop.
[B21: Base Flow Unit] Selectionofowunitsfor
theowratespan
This setting is linked with that of parameter C40.
Thisparameterselectstheowunitsfortheow
ratespan.(Incaseofmassow,thesettingof
density is also required. Read C46: Mass Flow
Density for more details.)
[B22: Base Time Unit] Selection of time units for
theowratespan
This setting is linked with that of parameter C41.
Thisparameterselectsthetimeunitsfortheow
rate span; however, if “m” has been selected for
theowrateunits,“/s”isautomaticallysetforthis
parameter.
[B23: Flow Span]Settingoftheowratespan
→ThissettingislinkedwiththatofparameterC42.
Thespancanbesetfortheforwardowrateinthe
range 0 to 32,000 (although this does not include 0).
The units set using B21/C40:BaseFlowUnit and
B22/C41:BaseTimeUnit will be displayed at this
time.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA ADMAG AXR and is the answer not in the manual?

YOKOGAWA ADMAG AXR Specifications

General IconGeneral
BrandYOKOGAWA
ModelADMAG AXR
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals