<11.  MAINTENANCE>
11-2
IM 01E30D01-01EN
11.2  Components Replacement
IMPORTANT
In case of explosion proof type, component 
replacement by other than authorized 
representative of Yokogawa is prohibited and 
willvoidthecertication.Ifnecessary,contact
Yokogawa.
WARNING
 
• Componentreplacementmustbecarriedout
by the trained personnel having knowledge 
of safety standard. No operator shall be 
permitted to perform any operations relating 
to component replacement.
• Componentreplacementandtheassociated
operations must be carried out by expert 
engineer or skilled personnel and not by 
operators.
• Beforeopeningthecover,itisimportantto
ensure that at least 5 minutes have passed 
since the power was turned off. Furthermore, 
opening of the cover must also be carried out 
by expert engineer or skilled personnel.
IMPORTANT
• Asarule,maintenanceofthisowmeter
should be implemented in a maintenance 
service shop where the necessary tools are 
provided.
• Theamplierassemblycontainssensitive
parts that may be damaged by static 
electricity. 
Excercise care so as not to directly touch 
the electronic parts or circuit patterns on 
the board, for example, by preventing static 
electricationbyusinggroundedwrist
straps when handing the assembly. Also 
take precautions such as placing a removed 
amplierassemblyintoabagwithan
antistatic coating.
11.2.1  Display Unit Replacement
CAUTION
When using this instrument for a long time in 
conditions of high temperature and humidity, the 
visibility of the display unit may deteriorate. In 
this case, it is necessary to replace the display 
unit.
(1)  Removing the Cover
(1)  Turn off the power.
(2)  The cover has the locking screws in front and 
behind it as shown in Figure 11.2.1 in case of 
explosion proof type. This is used for the cover-
locking system. Loosen cover locking screws 
clockwise using a hexagonal wrench (nominal 
size 3) to unlock the cover. (Upon shipment 
from the manufacturing plant, the cover is 
locked.) 
 Holdtheowmeterwithyourhandandremove
the cover by turning it in the direction of the 
arrow as shown below (both explosion proof 
and general purpose types).
F1103.ai
Cover locking 
screws
Front 
side
(1) (2)
Explosion proof type Both Explosion proof and 
General purpose types
Figure 11.2.1  Removing the Display Cover
(2)  Removing the Dispaly Unit
(1)  Hold the display unit with your hand and loosen 
the four mounting screws. Remove the display 
unit by pulling it in a straight, taking care not to 
damage it (Read Figure 11.2.2).