<Appendix1.  Safety Instrumented Systems Installation>
A1-1
IM 01E30D01-01EN
Appendix 1.  Safety Instrumented Systems 
Installation
WARNING
 
The contents of this appendix are cited from 
exida.comsafetymanualontheAXRseries
owmeterspecicallyobservedforthesafety
owmeterpurpose.WhenusingtheAXRfor
Safety Instrumented Systems (SIS) application, 
the instructions and procedures in this section 
must be strictly followed in order to preserve the 
owmeterforthatsafetylevel.
A1.1  Scope and Purpose
This section provides an overview of the user 
responsibilities for installation and operation of the 
AXRinordertomaintainthedesignedsafetylevel
for Safety Instrumented Systems (SIS) applications. 
Items that will be addressed are proof testing, repair 
and replacement of the transmitter, reliability data, 
lifetime, environmental and application limits, and 
parameter settings.
A1.2  Using the AXR for an SIS 
Application
A1.2.1  Safety Accuracy
TheAXRhasaspeciedsafetyaccuracyof2%.
This means that the internal component failures are 
listed in the device failure rate if they will cause an 
error of 2% or greater.
A1.2.2  Diagnostic Response Time
TheAXRwillreportaninternalfailurewithin8
seconds of the fault occurrence.
A1.2.3 Setup
Duringinstallationtheowmetermustbesetup
with engineering units parameters. This is typically 
done with a handheld terminal. These parameters 
mustbeveriedduringtheinstallationtoinsure
thatthecorrectparametersareintheowmeter.
Engineeringrangeparameterscanbeveriedby
reading these parameters from the optional local 
display or by checking actual calibration of the 
owmeter.
Thecalibrationoftheowmetermustbeperformed
after parameters are set.
A1.2.4  Required Parameter Settings
The following parameters need to be set in order to 
maintain the designed safety integrity.
Table A1.2.1  Required Parameter Settings
Item Description
Burnout direction 
switch
To specify if the output should go 
21.6 mA or higher or 3.2 mA or lower 
upon detection of an internal failure.
Write protection 
switch
The write function should be 
disabled.
A1.2.5  Proof Testing
The objective of proof testing is to detect failures 
withintheowmeterthatarenotdetectedbythe
diagnosticsoftheowmeter.Ofmainconcern
are undetected failures that prevent the safety 
instrumented function from performing its intended 
function. Read Table A1.2.2 for proof testing 
method.
The frequency of the proof tests (or the proof 
test interval) is to be determined in the reliability 
calculations for the safety instrumented functions 
forwhichtheAXRisapplied.Theactualprooftests
must be performed more frequently or as frequently 
asspeciedinthecalculationinordertomaintain
required safety integrity of the safety instrumented 
function.
Thefollowingtestsneedtobespecicallyexecuted
when a proof test is performed. The results of 
the proof test need to be documented and this 
documentation should be part of a plant safety 
management system. Failures that are detected 
should be reported to Yokogawa.
The personnel performing the proof test of the 
owmetershouldbetrainedinSISoperations
includingbypassprocedures,AXRowmeter
maintenance, and company management of 
change procedures.