6-31
IM AQ6370D-01EN
Waveform Display
6
Displaying Wavelength Line Markers
1.
Press the MARKER. The soft key menu for marker settings appears.
2.
Press the MORE 1/3 soft key.
3.
Press the LINE MARKER 1 OFF/ON or LINE MARKER 2 OFF/ON soft key to
turn the function ON. Line marker values are displayed in the upper left of the
waveform area.
Note
• Wavelength line markers cannot be displayed if the active trace measurement span is 0.000
nm.
• When both wavelength line markers 1 and 2 are displayed, the wavelength difference (L2–
L1) is shown below the marker values.
Moving Line Markers
4.
Refer to the following and move the line marker.
Direction Moving Procedure
Move to right Turn the rotary knob to the right.
Press the UP arrow key.
Move to left Turn the rotary knob to the left.
Press the DOWN arrow key.
Clearing Line Markers
1.
Press the MARKER. The soft key menu for marker settings appears.
2.
Press the MORE 1/3 soft key.
3.
Press the LINE MARKER 1 OFF/ON or LINE MARKER 2 OFF/ON soft key to
turn the function OFF.
6.9 Marker Display