8-21
IM AQ6370D-01EN
Saving/Loading Data
1
2
3
4
5
6
7
8
9
10
11
App
Index
Waveform Data
The measured waveform data is stored as sets of wavelength (in nm) and level. The
number of sets equals the number of measurement samples.
Waveforms measured in frequency mode are stored using their frequencies (THz).
Level values are stored as log values if the vertical axis scale is LOG, and as linear
values if the vertical axis scale is linear.
(For LOG Scale)
[TRACE DATA] Header indicating the start of the trace data
****.****, ±***.***(CR)(LF) Wavelength value and level value (LOG) of first point
****.****, ±***.***(CR)(LF) Wavelength value and level value (LOG) of second point
:
****.****, ±***.***(CR)(LF) Wavelength value and level value (LOG) of final point
(For LINEAR Scale)
[TRACE DATA] Header indicating the start of the trace data
****.****,*.***E±***(CR)(LF) Wavelength value and level value (LINEAR) of first point
****.****,*.***E±***(CR)(LF) Wavelength value and level value (LINEAR) of second point
:
****.****,*.***E±***(CR)(LF) Wavelength value and level value (LINEAR) of final point
8.3 Saving/Loading Displayed Data

Table of Contents

Other manuals for YOKOGAWA AQ6370D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA AQ6370D and is the answer not in the manual?

YOKOGAWA AQ6370D Specifications

General IconGeneral
BrandYOKOGAWA
ModelAQ6370D
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals