Special measures
8-7CenterMax Operating Instructions61211-1020202
• Push against the marked spot on the cover and move the cross-
beam back into the measuring volume.
Depending on the crossbeam position, you have to push against
the crossbeam from the left- or right-hand side.
Important!
You may only push against the crossbeam at the drive end.
• Do not push against the probe head and the crossbeam at the
drive side.
The probe head, quill or crossbeam may be damaged.
z direction: You must release the quill fall brake before moving the
probe head back into the measuring volume. Proceed as follows:
• Move the crossbeam in the y direction to the front area of the
measuring volume. Always observe the following: ➤ page 8-6.
Move the crossbeam forward far enough so that you can grasp the
cover with your hand.
The right-hand side of the cover (1) is provided with an opening. You
can access the switch for deactivating the fall brake through this
opening.
!