Section 10 Troubleshooting and Diagnostics
Tower/Loader Fault 5: communication failure/BL
A communication failure occurred between the Analyzer and the module that
controls the tower and loader.
– - Probable Cause – - Corrective Action
1. A communication timing or
handshake error has occurred.
1. Reboot the System.
NOTE: If you wish to use the
Sample Loader, reset the
racks before rebooting the
System.
2. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
Tower/Loader Fault 16: direct command parameter error/BL
The Analyzer command given to the module that controls the tower and loader
is invalid.
– - Probable Cause – - Corrective Action
1. A software error has occurred. 1. Reboot the System.
NOTE: If you wish to use the
Sample Loader, reset the
racks before rebooting the
System.
2. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
Table 10.2 General Fault Conditions (Continued)