6243/44 DC Voltage Current Source/Monitor Operation Manual
4.4.1 Staircase Sweep
4-15
(2) Changing parameters for sweep measurement
Parameters for sweep measurement can only be changed in Standby status. However, the following
items can be changed during sweep stop in the Operate status.
Time parameters
•Hold time
Source Delay time
Measure Delay time
• Period
Pulse width
Auto Range Delay time
Sweep Trigger Mode
Start address and Stop address of Random Sweep
(Within the range of the Start address and Stop address specified before operation)
Repeat count
Reverse Mode ON/OFF
•Bias Value
4.4.1 Staircase Sweep
Staircase sweeps, linear sweeps or log sweeps can be selected for the sweep type and one-way sweep or
round-trip sweep can be selected with Reverse ON/OFF.
Table 4-5 Sweep Types for Staircase Sweep
Sweep type Output waveform Number of steps, No. K output value
Linear sweep
Reverse-OFF
No. K output= Start value + (K-1) (Step value)
Number of steps= | Stop value - Start value | /Step value + 1
(Maximum 5000)
Linear sweep
Reverse-ON
Log sweep
Reverse-OFF
No. K output = (Start value) 10 K/n
n: Number of divisions per decade
Start value, Stop value0
| Start value | | Stop value |
Start value polarity = Stop value polarity
Log sweep
Reverse-ON