6243/44 DC Voltage Current Source/Monitor Operation Manual
3.3.7 MENU key (Setting Parameters)
3-12
3.3.7.5 COMPARATOR (Comparison Operation Parameters)
Selecting COMPARATOR and pressing the key enables selection of parameters related to comparator
calculation.
Compare Toggles comparator calculation between ON and OFF.
ON: The following comparator calculations are per-
formed.
Upper Data  Measurement result; HI
Lower Data ï‚£ Measurement result ï‚£ï€ Upper Data;
GO
Measurement result  Lower Data; LO
OFF: Comparator calculation is set to OFF.
Upper Data Enables the upper limit value setting for comparator calculation.
Lower Data Enables the lower limit value setting for comparator calculation.
Result Buzzer Selects the condition on which the buzzer sounds in accordance
with the results of comparator calculation.
HI: The buzzer sounds when the result of the compar-
ator calculation is HI.
GO: The buzzer sounds when the result of the compar-
ator calculation is GO.
LO: The buzzer sounds when the result of the compar-
ator calculation is LO.
OFF: The buzzer is OFF.
3.3.7.6 EXT.SIGNAL (External Control Signal Parameters)
Selecting EXT.SIGNAL and pressing the key enables selection of parameters related to the external
control signal input on the rear panel.
Operate Signal Selects the function of the INTERLOCK OPERATE IN/OUT
signal input on the rear panel.
Operate OFF In:
Sets Standby when the input signal rises from Lo
to Hi.
Operate ON/OFF In:
Sets Standby when the input signal rises from Lo
to Hi, and sets Operate when the signal falls from
Hi to Lo.