EasyManua.ls Logo

ADCMT 6244 - Range and Data Differences in Accordance with Range Specification (Example: Showing DC Sweep with 6243)

Default Icon
296 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6243/44 DC Voltage Current Source/Monitor Operation Manual
4.5.3 Source Range
4-24
Set values and decided ranges for auto range
Taking the example of DC sweep, Table 4-9 shows how the source range and source data become re-
sults of the range specification for the same source value.
Setting value
Specified
range
6243
Voltage
source
0 V |V| 320 mV
320 mV |V| 3.2 V
3.2 V |V| 32 V
32 V |V| 110 V
320 mV
3.2 V
32 V
110 V
Current
source
0 A | I | 32 A
32 A | I | 320 A
320 A | I | 3.2 mA
3.2 mA | I | 32 mA
32 mA | I | 320 mA
320 mA | I | 2 A
32 A
320 A
3.2 mA
32 mA
320 mA
2 A
6244
Voltage
source
0 V |V| 320 mV
320 mV |V| 3.2 V
3.2 V |V| 20 V
320 mV
3.2 V
20 V
Current
source
0 A | I
| 320 A
320 A | I | 3.2 mA
3.2 mA | I | 32 mA
32 mA | I | 320 mA
320 mA | I | 3.2 A
3.2 A | I | 10 A
320 A
3.2 mA
32 mA
320 mA
3.2 A
10 A
Table 4-9 Range and Data Differences in Accordance With Range Specification
(Example: Showing DC Sweep With 6243)
Source setting
Fixed range
(110 V range fixed)
Auto range
Parameter
Setting
value
RangeDataRangeData
Bias value
Start value
Stop value
0.1 V
2 V
80 V
110 V
110 V
110 V
0.10 V
2.00 V
80.00 V
320 mV
3.2 V
110 V
100.00 mV
2.0000 V
80.00 V
110 V range is selected for all.
320 mV range is selected for Bias values.
3.2 V range is selected for Start values.
110 V range is selected for Stop values.
320 mV range to 110 V range is used during sweep.