6243/44 DC Voltage Current Source/Monitor Operation Manual
9.1.4 Setting Time
9-10
9.1.4 Setting Time
Minimum pulse width: 1 ms
Minimum step (repeat) time: Source range fixed, free run or internal trigger mode, source delay time
10 s
Measurement range fixed, integration time 500 s, and measure delay time 300 s when measurement
is ON.
Integration time: 500 s/1 ms/10 ms/1 PLC/10 PLC/100 PLC
Source delay time:
Period (pulse cycle):
Pulse width:
Measure delay time:
Measurement
Memory mode
Minimum step time
OFF — 2 ms
ON
BURST 4 ms
NORMAL
10 ms
OFF
Setting range
Resolution
Accuracy
10 s to 600.00 ms 10 s
0.1% + 30 s)600.1 ms to 6000.0 ms 100 s
6001 ms to 60000 ms 1 ms
Setting range Resolution Accuracy
2 ms to 600.00 ms 10 s
0.1% + 30 s)600.1 ms to 6000.0 ms 100 s
6001 ms to 60000 ms 1 ms
Setting range Resolution Accuracy
1 ms to 600.00 ms 10 s
0.1% + 30 s)600.1 ms to 6000.0 ms 100 s
6001 ms to 60000 ms 1 ms
Setting range Resolution Accuracy
300 s to 600.00 ms 10 s
0.1% + 30 s)600.1 ms to 6000.0 ms 100 s
6001 ms to 60000 ms 1 ms