6243/44 DC Voltage Current Source/Monitor Operation Manual
5.2.10 GPIB Code List
5-30
5.2.10.2 TER? Command Response
Table 5-6 shows the responses and their meanings when the results of self-tests are recalled using the TER?
command. Figures are added for data where multiple reasons for the error exist.
Response: a, b, c, d
Example: When the response is 0,24,0,0
Indicates that self-test errors occurred at the two positions AD 3.2 V ZERO and AD 320 mV ZERO
Table 5-6 TER? Command Response (1 of 2)
Register Data Error message
a
0
1
2
4
8
16
Pass (No error)
RAM Read/Write
Logic-Panel Communication
Logic-Analog Communication
CAL Data Check Sum
Parameter Check Sum
b
0
1
2
4
8
16
Pass (No error)
AD Ratio IR1/IR2
AD Ratio IR2/IR3
AD Ratio IR3/IR4
AD 3.2V ZERO
AD 320mV ZERO
c
0
1
2
4
8
16
32
64
128
256
512
1024
2048
4096
8192
16384
32768
Pass (No error)
VSVM 320mV +ZERO
VSVM 320mV -ZERO
VSVM 320mV +FS
VSVM 320mV -FS
VSVM 3.2V +ZERO
VSVM 3.2V -ZERO
VSVM 3.2V +FS
VSVM 3.2V -FS
VSVM 32V +ZERO /VSVM 20V + ZERO
VSVM 32V -ZERO /VSVM 20V - ZERO
VSVM 32V +FS /VSVM 20V + FS
VSVM 32V -FS /VSVM 20V - FS
VSVM 110V +ZERO
VSVM 110V -ZERO
VSVM 110V +FS
VSVM 110V -FS

Table of Contents