9.2.5 General Specifications
9-23
9.2.5 General Specifications
Operating environment conditions: Ambient temperature 0C to 50C. Relative humidity: 85% or
lower (with no condensation)
However, the operating temperature is 0C to 40C in the follow-
ing output ranges:
When 0 V Vo 7 V: Io Vo - 10 [A]
When -7 V Vo 0 V: Io Vo + 1 0 [ A ]
Vo: Voltage between output terminals [V]
Io: Output current [A]
Storage environment conditions: Ambient temperature -25C to +70C. Relative humidity:
85% or lower (with no condensation)
Warm-up: 60 minutes or longer (until specified accuracy is reached)
Display: 5 7 dot matrix fluorescent character display tube
Power supply: AC power, 100 V, 120 V, 220 V, 240 V (user switchable)
Line frequency: 50 Hz/60 Hz
Power consumption: 400 VA or less
External dimensions: Approximately 212 (W) 177 (H) 500 (D) mm
Mass: 15 kg or lower
Option No. Standard opt 32 opt 42 opt 44
Power supply
voltage
100 V 120 V 220 V 240 V