6243/44 DC Voltage Current Source/Monitor Operation Manual
4.5.2 Source value Restrictions
4-22
(2) Mode change examples
When set as follows: Source Value: +1.0000 V. Limiter Value: 0500.0 mA. Bias Value:
+0.0000 V.
Source Mode Source Value Limiter Value
VS: +1.0000 V L: 0500.0 mA
VS: +1.0000 V L: 0500.0 mA
VS: +0.0000 V L: 0500.0 mA
VS: +0.0000 V L: 0500.0 mA
4.5.2 Source value Restrictions
(1) Output range
The setting range for the source value is restricted by the limiter (compliance) value as shown in the
following table. If set out of this range, an error is generated.
Table 4-7 Source Value Setting Range (6243)
Function Limiter setting Range
Voltage source
0.3 AI
L
0.5 A
0.5 A I
L
1 A
1 A I
L
2 A
0 to 110 V
0 to 64 V
0 to 32 V
Current source
3 mVV
L
32 V
32 V V
L
64 V
64 V V
L
110 V
0 to 2 A
0 to 1 A
0 to 0.5 A
Table 4-8 Source Value Setting Range (6244)
Function Limiter setting Range
Voltage source
3 AI
L
4 A
4 A I
L
10 A
0 to 20 V
0 to 7 V
Current source
3 mVV
L
7 V
7 V V
L
20 V
0 to 10 A
0 to 4 A
DC
PLS
SWP
PSW